Specifications: 59 MAX Infrared Thermometer
Specifications | ||
Temperature Range | -30°C to 350°C (-22°F to 662°F) | |
Accuracy (Calibration geometry with ambient temperature 23°C ±2°C) | ≥ 0°C | ±2.0°C or ±2.0% of reading, whichever is greater |
≥ -10°C to <0°C | ±2.0°C | |
< -10°C | ±3.0°C | |
≥ 32°F | ±4.0°F or ±2.0% of reading, whichever is greater | |
≥ 14°F to <32°F | ±4.0°F | |
< 14°F | ±6.0°F | |
Response Time (95%) | < 500 ms (95% of reading) | |
Spectral Response | 8 mm to 14 mm | |
Emissivity | 0.10 to 1.00 | |
Distance to Spot Ratio | 8:1 (calculated at 90% energy) | |
Display Resolution | 0.1°C (0.2°F) | |
Repeatability (% of reading) | ±1.0% of reading or ±1.0°C (±2.0°F), whichever is greater |
|
Power | 1 AA IEC LR06 Battery | |
Battery Life | 12 hours with laser and backlight on | |
Weight | 220 g (7.76 oz) | |
Size | 156 x 80 x 50 mm (6.14 x 3.15 x 2 in) | |
Operating Temperature | 0°C to 50°C (32°F to 122°F) | |
Storage Temperature | -20°C to +60°C (-4°F to 140°F), (without battery) | |
Operating Humidity | 10% to 90% RH non-condensing @ 30°C (86°F) | |
Operating Altitude | 2000 meters above mean sea level | |
Storage Altitude | 12,000 meters above mean sea level | |
Drop Test | 1 m | |
Ingress Protection Rating | IP40 per IEC 60529 | |
Vibration and Shock | IEC 68-2-6 2.5 g, 10 to 200 Hz, IEC 68-2-27, 50 g, 11 ms | |
Compliance | EN/IEC 61010-1 | |
Laser Safety | FDA and EN 60825-1 Class II | |
Electromagnetic Compatibility | 61326-1 EN 61326-2 |